Calibrated equipment

Measuring receivers (EMI receivers, spectrum analyzers)

Calibration method - according to EN 55016-1-1 standard.

If you do not need the full calibration scope please provide:

  • Detectors (QP, Peak, Average/CISPR Average, RMS),
  • Number of measurement inputs,
  • Frequency scope for each of the measurement inputs.

Optionally: statement of conformity of the results with the requirements of the EN 55016-1-1 standard.

Pulse generators

Calibration method – according to EN 55016-1-1 standard.

Optionally: statement of conformity of the results with the requirements of the EN 55016-1-1 standard .

Discontinuous disturbance analyzers

Calibration method - according to EN 55016-1-1 standard.

Optionally: statement of conformity of the results with the requirements of the EN 55016-1-1 standard.

Artificial networks / ISNs

Calibration method - according to EN 55016-1-2 standard.

Calibration scope - in accordance with the technical documentation of the device and the applicable standard.

If the network operates in different modes (e.g. limiter/filter/internal terminator), please specify the settings in which the calibration is to be performed.

Optionally: statement of conformity of the results with the requirements of EN 55016-1-2, ISO 7637-2 and EN 55025 standards.

Absorption clamps

Calibration method - according to EN 55016-1-3 standard.

Please specify in the inquiry, whether the absorption clamps are to be calibrated with an attenuator and a cable.

Antennas

Calibration method - according to EN 55016-1-6, SAE ARP 958, or ANSI standard.

Please provide the following:

  • Calibration standard,
  • Measurement distance,
  • Frequency scope and measurement step,
  • Polarizations (horizontal, vertical),
  • Modes (if the antenna has any).

Voltage probes

Calibration method - according to EN 55016-1-2 standard.

Optionally: statement of conformity of the results with the manufacturer's technical documentation.

Current probes/ Pearson probes

Calibration method - according to EN 55016-1-2 standard.

Calibration scope for the current probe – in accordance with its technical documentation.

Power meters and probes

Calibration method – in compliance with the Laboratory’s in-house, accredited method.

Basic calibration is performed as a function of frequency at: 10 dBm, 0 dBm, -15 dBm. Please let us know if testing at other levels is required.

CDNs

Calibration method - according to EN 55016-1-2 standard.

Optionally: statement of conformity of the results with the requirements of the EN 61000-4-6 standard.

ESD generators

Calibration method - according to EN 61000-4-2 or ISO 10605 standard.

Optionally: statement of conformity of the results with the requirements of EN 61000-4-2 or ISO 10605 standards.

EFT/B generators with additional devices

Calibration method - according to EN 61000-4-4, ISO 7637-2 or ISO 7637-3 standard.

Optionally: statement of conformity of the results with the requirements of EN 61000-4-4, ISO 7637-2 or ISO 7637-3 standards.

SURGE generators with additional devices

Calibration method - according to EN 61000-4-5 standard.

Optionally: statement of conformity of the results with the requirements of the EN 61000-4-5 standard.

Passive elements

Calibration method – in compliance with the Laboratory’s in-house, accredited calibration method focusing on attenuation and standing wave ratio.

Ring wave generators

Calibration method - according to EN 61000-4-12 standard.

Optionally: statement of conformity of the results with the requirements of the EN 61000-4-12 standard.

Damped oscillatory wave generators

Calibration method - according to EN 61000-4-18 standard.

Optionally: statement of conformity of the results with the requirements of the EN 61000-4-18 standard.

Generators for voltage dips, short interruptions and voltage variations

Calibration method - according to PN- EN 61000-4-11 standard.

Optionally: statement of conformity of the results with the requirements of the PN- EN 61000-4-11 standard.

Calibration lead time:

Approximate calibration lead time per one device. For higher device quantities, the lead time is determined on a case-by-case basis.

Standard – 5-10 business days.

Express – From 3 business days. Please inquire for details.

Without any prior arrangements concerning the completion date – up to 15 business days.

The lead times specified above do not apply to antennas requiring measurements performed at an Open Area Test Site (OATS).